DeVilbiss 525 Series Service Manual Page 17

  • Download
  • Add to my manuals
  • Print
  • Page
    / 44
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 16
LT-2023
17
MOLECULAR SIEVE BEDS
The two molecular sieve beds alternately remove the nitrogen from the
air passing through them and provide the patient with a constant supply of
oxygen.
The efficiency of the molecular sieve material will be impaired if it
becomes contaminated by moisture. Contamination causes the molecular
sieve material to lose its nitrogen adsorbing properties resulting in a
decrease in oxygen concentration. The unit should run for a minimum of
20 minutes before turning "Off" to prevent problems associated with
moisture contamination in the system.
To test the sieve beds:
1. Remove the 1/16" plugged piece of tubing from the top of each sieve
bed.
2. Connect the pressure-vacuum gauges to the sieve bed test points
(Figure 9) in order to observe unit cycling and bed pressures. Refer
to the section on Normal Operating Sequence.
3. If it is determined that the valve did not shift, refer to the section on
Rotary Valve testing. However, if the unit is cycling properly allow
unit to run while observing the sieve bed pressures.
4. After 20 minutes of operation, check the oxygen concentration
levels. Low oxygen concentration and high pressures indicate
contaminated sieve beds.
NOTE– If the molecular sieve material is found to be no longer
effective,rstlocatethesourceofthemalfunctionorcause
(suchasleaks)forcontaminationandtakecorrectiveaction.
To replace the molecular sieve beds (Figs. 5 & 6):
1. Make sure any contamination problem has been corrected before
replacing.
2. Make sure the unit is unplugged from the wall outlet.
3. Cut the plastic cable ties that secure the sieve beds to the internal
structure of the unit.
4. Remove the tubing from the fittings at the top of each sieve bed.
5. Remove the hose clamps and hose from the bottom of the sieve
beds.
6. Install new sieve beds in reverse order using new plastic cable ties.
Position the new beds so that the bed serial number label is at the
top of the unit.
NOTE– Make sure that the sealing caps remain on the new sieve beds
untiljustpriortoconnectinghosesandtubing.
7. Leak test all connections with a certified leak detection solution such
as Snoop® or equivalent (must not contain ethylene glycol). Apply
leak test solution to all fittings and hose connections with unit
running. If an air leak is present, the solution will bubble. All leaks
should be repaired before putting the unit back in service.
CAUTION– Do not apply leak test solution to any part of the rotary
valve or the main PC Board assembly.
POWER CORD
To replace the power cord - 115 volt units only
(Fig. 11 & 14):
1. Make sure the unit is unplugged from the wall outlet.
2. Remove the back cabinet.
3. Disconnect the power cord connector.
NOTE–Looseningorremovingthecompressormountingplatemay
make it easier to service the power cord.
4. Note wire colors and socket locations before removing wires.
5. Using a pair of duckbill pliers, squeeze the power cord strain relief
and pull it out of the base of the unit.
6. Insert a new power cord through the hole in the base of the unit and
secure with strain relief.
7. Insert sockets into connector housing and then reconnect the power
cord connector.
8. Replace back cabinet and secure with the six screws.
POWER SWITCH
To replace the power switch (Figure 10):
1. Make sure the unit is unplugged from the wall outlet.
2. Remove the front and back cabinets and loosen the bib.
3. Note the position of the wires and switch before removing the wires
from the switch terminals.
4. While squeezing the locking tabs on the sides of the switch, push the
switch out of the front of the bib.
5. Install the new switch in the correct orientation making sure that it
locks into position.
6. Reconnect the wires to the switch terminals.
PRESSURE REGULATOR
The pressure regulator stabilizes the flow of oxygen to the patient and
establishes back pressure on the system. It is preset at 8.5 psi (58.6 kPa)
and should not have to be adjusted in the field.
To test the pressure regulator:
1. Turn the unit On.”
2. Set the flow meter at 2 lpm.
3. Use a pressure-vacuum gauge (part #PVO2D-601) and a fitting
suitable to fit on the oxygen outlet or on a short piece of tubing
connected to the outlet.
4. If the pressure-vacuum gauge reads anything other than 8.5 ± .5 psi
(58.6 ± 5.9 kPa) with gauge outlet blocked, adjustment to the
pressure regulator may be required. If so, call DeVilbiss Service
Department at 1-800-338-1988 (814-443-4881).
NOTE–Makesurenoleaksexistbeforeadjustingthepressure
regulatorbyusingacertiedleakdetectionsolutionsuchas
Snoop®orequivalent(mustnotcontainethyleneglycol).
CAUTION– Do not apply leak test solution to any part of the rotary
valve or the main PC Board assembly.
A malfunction in the pressure regulator will cause either a loss or
fluctuation in the oxygen flow which will be seen on the flow meter or a
decrease in oxygen concentration.
COMPONENT TESTING, REPAIR, AND REPLACEMENT
Brown Wire (or 2 Black Wires)
2 Black Wires (or 1 Brown Wire)
*Red Wire
*Red Wire
* Either red wire can be connected to
either left hand terminal of the switch
Switch Detail
Page view 16
1 2 ... 12 13 14 15 16 17 18 19 20 21 22 ... 43 44

Comments to this Manuals

No comments