DeVilbiss 525 Series Service Manual Page 16

  • Download
  • Add to my manuals
  • Print
  • Page
    / 44
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 15
LT-2023
16
8. Place valve plates on the compressor so that heads of reed valve
screws are aligned with the indentation in top of pistons.
9. Install the compressor heads so that the holes in the heads are
aligned with the holes in the compressor housing.
10. Secure compressor heads with the screws.
To replace the compressor:
CAUTION– The525DSconcentratorsuseaGSEcompressorwith
a22mfdcapacitor.The525KSconcentratorsusea
GSEcompressorwitha10mfdcapacitor.If
replacement is necessary, be sure the correct capacitor
is installed.
1. Inspect the motor mounts. Replace if damaged. Attach mounts to
compressor feet.
2. Inspect the capacitor to determine if replacement is necessary. If
capacitor is wrong value for compressor or replacement is desired,
refer to Capacitor section.
3. Reconnect hose to the compressor intake fitting.
4. Position compressor on the base of the unit so that the studs on the
motor mounts protrude through the holes in the base of the unit.
5. Secure motor mounts with hex nuts.
6. Reconnect hoses to the fittings at compressor exhaust and filter.
7. Reconnect the compressor electrical connector.
COOLING FAN
The cooling fan provides a constant air flow to cool the compressor. The
cooling fan is located in the bottom of the unit below the compressor
(Figure 14).
A defective cooling fan may cause the compressor’s internal thermo-
protective device to activate and shut the compressor off. Should this
condition occur, the compressor will require several minutes for the
thermo-protective device to reset.
If the cooling fan is defective, it must be replaced:
1. Make sure the unit is unplugged from the wall outlet.
2. Remove the front and back cabinets.
3. Remove the compressor.
4. Disconnect the cooling fan terminals.
5. Note the position of the fan and fan guard before removing the four
retaining screws that secure the fan to the base of the unit.
6. Remove the defective fan and secure the replacement fan in position
with the four retaining screws.
NOTE–Wheninstallingthefan,besuretheairowdirectionalarrowon
the side of the fan is directed away from the compressor and
fan guard is reinstalled properly.
7. Reconnect the electrical connector.
8. Reinstall the compressor.
FINAL CHECK VALVE
This check valve is located between the final bacteria filter and the oxygen
outlet fitting. This check valve allows oxygen to flow only out of the unit.
When the unit is turned off and oxygen flow stops, the check valve closes
to prevent water from being drawn into the unit.
A defective final check valve may allow water to be drawn in from the
humidifier bottle when the unit is turned off. This may occlude the final
bacteria filter and/or the flow meter causing a restriction of flow and
making it difficult to adjust the flow rate.
To replace the final check valve (Figure 10):
1. Make sure the unit is unplugged from the wall outlet.
2. Remove the front and back cabinets and loosen or remove the bib.
3. Remove the hose from the outlet side of the final bacteria filter.
4. Remove the two screws from the back of the oxygen outlet fitting
assembly and remove the assembly.
5. Remove the hose from each end of the final check valve.
6. Attach the hoses to a new check valve. Make sure that the flat side
of the check valve is directed toward the oxygen outlet fitting.
7. Replace the outlet fitting assembly and connect the hose to the filter.
8. Replace the bib and front and back cabinets.
FLOW METER
The pressure-compensated flow meter has an accuracy level of ±5% at
full scale (exception: +0%,-5% at 5 lpm). The flow meter on the
DeVilbiss Oxygen Concentrator is designed for use at 8.5 psi (58.6 kPa) at
flow rates up to 5 lpm.
To check for leaks in the flow meter tubing:
1. Check for leaks using a certified leak detection solution such as
Snoop® or equivalent (must not contain ethylene glycol).
2. Apply leak test solution to all fittings and hose connections with the
unit running.
CAUTION– Do not apply leak test solution to any part of the rotary
valve or the main PC Board assembly.
3. f an air leak is present, the solution will bubble. All leaks should be
repaired before putting the concentrator back in service.
Electric Shock Hazard. Use caution when leak testing near
electrical connections.
To replace the flow meter (Figure 10):
1. Make sure the unit is unplugged from the wall outlet.
2. Remove the front and back cabinets.
3. From behind the bib remove the 2 hoses from the flow meter.
4. While squeezing tabs on flow meter brackets, push the flow meter
through the bib.
5. Install new flow meter in bib and reconnect hoses.
HOUR METER
To replace the hour meter (Figure 9):
1. Make sure the unit is unplugged from the wall outlet.
2. Remove the front and back cabinets and loosen the bib.
3. Disconnect the hour meter connector from the PC board.
4. Remove the meter by carefully inserting a small flat screwdriver
under the outer edge of meter and prying upward.
5. Install a new hour meter by applying downward pressure until it
snaps into position.
6. Connect the hour meter to the PC board.
CAUTION– DonotapplyanyforceorexthePCBoardwhen
connecting or disconnecting electronic or pneumatic
components. Damage to the electronic assembly is
possible.
COMPONENT TESTING, REPAIR, AND REPLACEMENT
WARNING
Page view 15
1 2 ... 11 12 13 14 15 16 17 18 19 20 21 ... 43 44

Comments to this Manuals

No comments