For Product Numbers:306 SeriesSERVICE MANUALf o r q uali fi ed s e r v i c e p e r s o n n e l o n l yDANGER – NO SMOKING CAUTION-Federal (U.S.A.
10LT-2044ENSYMPTOM POSSIBLE CAUSE REMEDYDC Adapter attached. Unit operating from Battery or not operating. External Power light is not illuminated1.
11LT-2044ENTypical New Battery Operation TimeFlow Rate Continuous Flow PulseDose (20 BPM)1.0 4.0 hours 5.4 hours2.0 2.4 hours 4.7 hours3.0
12LT-2044ENCOMPONENT TESTING, REPAIR, AND REPLACEMENTAt this time, DeVilbiss recommends limited replacement of components and no testing or field repai
13LT-2044ENre a d a l l o f t h e s t e p s i n v o l v e d b e f o r e b e G i n n i n G a n y o f t h e p r o c e d u r e s i n t h i s
14LT-2044ENMotor MountsTo replace the motor mounts:1. Remove covers from the unit.2. Lay the unit onto a flat surface with the compressor facing up
15LT-2044ENORDERING PARTSDeVilbiss Customer Service Contact InformationCustomer Service (USA) 800-338-1988 / 814-443-4881, [email protected]•Can
16LT-2044ENToolsOxygen Analyzer O2ANAPressure Test Assembly 303DZ-637Pressure Vacuum Gauge PVO2D-601Service Kit
17LT-2044ENCAUTION– When moving the iGo from an extreme environment, allow time for the iGo to acclimate to the recommended operating environment. Ope
18LT-2044ENFIGURES, DIAGRAMS AND VIEWSIndexDiagramsFigure 1 - Pneumatic DiagramFigure 2 - Wiring DiagramExterior ViewsFigure 3 - FrontFigure 4 - RearF
19LT-2044ENFigure 1 - Pneumatic DiagramfIGuRes, DIaGRams anD vIeWsLT-2044 Rev A 306 iGo SM.indd 19 11/6/09 6:23:23 AM
2LT-2044ENtable of contentsIMPORTANT SAFEGUARDS ...
20LT-2044ENFigure 2 - Wiring DiagramfIGuRes, DIaGRams anD vIeWsLT-2044 Rev A 306 iGo SM.indd 20 11/6/09 6:23:23 AM
21LT-2044ENFigure 3 - FrontOxygen OutletControl PanelFront CoverCover ScrewsBasefIGuRes, DIaGRams anD vIeWsLT-2044 Rev A 306 iGo SM.indd 21 11/6/0
22LT-2044ENFigure 4 - RearBaseBatteryRear CoverBattery LatchHandleAir Filter(in handle cavity)Cart ConnectionfIGuRes, DIaGRams anD vIeWsLT-2044 Rev
23LT-2044ENFigure 5 - SideFront CoverBaseHandle (one on each side)Power InputRear CoverExhaust VentsfIGuRes, DIaGRams anD vIeWsLT-2044 Rev A 306 iGo
24LT-2044ENFigure 6 - BottomFigure 6B - Inset Close UpFigure 6A - APP Tubing InsideRating LabelAtmospheric Pressure Port (see close up below)Atmospher
25LT-2044ENFigure 7 - Battery BayBattery ContactsBattery SlotsCommunication PortBattery BumpersCover ScrewsfIGuRes, DIaGRams anD vIeWsLT-2044 Rev A
26LT-2044ENFigure 8 - Top with Rear Cover RemovedFront CoverRibbon Connector CableAccumulator TankFinal Bacteria FilterIntake Silencer/FilterfIGuRes,
27LT-2044ENFigure 9 - Inside Front CoverRibbon Connector CableFoam1/8" TubingOxygen OutletfIGuRes, DIaGRams anD vIeWsLT-2044 Rev A 306 iGo SM.i
28LT-2044ENFigure 10 - Rear with Cover RemovedFour-Way ValveBaseSieve BedSieve BedRegulatorPC BoardfIGuRes, DIaGRams anD vIeWsLT-2044 Rev A 306 iGo
29LT-2044ENFigure 11 - Front with Cover RemovedAccumulator TankIntake Silencer/Filter (306DS-616)Pressure Relief ValveExhaust MufflerCompressorMotor M
3LT-2044ENIMPORTANT SAFEGUARDSWhen using electrical products, especially when children are present, basic safety precautions should always be followed
30LT-2044ENFigure 12 - Motor BoardfIGuRes, DIaGRams anD vIeWsLT-2044 Rev A 306 iGo SM.indd 30 11/6/09 6:23:33 AM
31LT-2044ENFigure 13 - PC BoardPressure SwitchOSDfIGuRes, DIaGRams anD vIeWsLT-2044 Rev A 306 iGo SM.indd 31 11/6/09 6:23:34 AM
DeVilbiss Healthcare100 DeVilbiss Drive Somerset, PA 15501 USA(814) 443-4881(800) 338-1988In Canada (800) 263-3390© 2009 DeVilbiss Healthcare 07.09LT
4LT-2044ENGENERAL INFORMATIONIntroductionRefer to the instruction guide (A-306 or D-306) for set up and operating instructions. This service manual is
5LT-2044ENSet Up & Operating InstructionsRefer to the instruction guide (A-306 or D-306) for set up and operating instructions. An overview is bel
6LT-2044ENMAINTENANCEAlert SystemThe iGo alert system is comprised of both visible and audible alerts which signal a malfunction in operation. The ale
7LT-2044ENPeriodic Provider MaintenanceSummary3 years: Inspect intake silencer/filter—replace as needed Check oxygen concentration.6 years: In
8LT-2044EN 1) If the battery is installed on the iGo, clean it while it is installed using a damp cloth with a mild household cleaner and wipe dry
9LT-2044ENSYMPTOM POSSIBLE CAUSE REMEDYUnit not operating. External Power light flashing with audible alert. Power Fail alert activated. 1. AC power