SYMPTOM POSSIBLE CAUSE REMEDY
Unit not operating. External
Power light flashing with
audible alert. Power Fail alert
activated.
1. AC power cord not properly inserted into
wall outlet or DC adapter not attached.
2. Charged battery not installed or defective
battery
3. No power at wall outlet.
4. Faulty AC Adapter.
5. Faulty DC accessory power port outlet.
6. Faulty DC Adapter.
1. Check power cord connection at the wall outlet and adapter
connection to iGo.
2. Install charged battery or contact your provider for a
replacement battery.
3. Check the home circuit breaker and reset if necessary. Use a
different wall outlet if the situation occurs again.
4. Contact your DeVilbiss provider.
5. Check automotive fuse.
6. Contact your DeVilbiss provider.
If the above remedies do not work, contact DeVilbiss.
Unit set for PulseDose mode:
No Breath Detected light
flashing and audible alert
sounding.
OR
Unit changed to Continuous
Flow automatically.
1. Cannula is not adjusted properly.
2. Obstructed cannula or oxygen tubing.
3. Tubing/cannula too long.
4. Humidifier attached.
5. Low flow cannula being used.
6. Atmospheric Pressure Port obstructed.
7. Tubing inside unit is kinked.
8. Patient not candidate for PulseDose delivery.
1. Check all cannula connections to make sure they are tight
and adjust the cannula to fit comfortably. Ensure tubing is not
kinked.
2. Detach cannula. If proper flow is restored, clean or replace if
necessary. Disconnect the oxygen tubing at the oxygen
outlet. If proper flow is restored, check oxygen tubing for
obstructions or kinks. Replace if necessary.
3. Replace with shorter tubing/cannula.
4. Remove humidifier. (PulseDose does not operate with
humidifier)
5. Replace with standard cannula.
6. Remove obstruction.
7. Unkink tubing.
8. Use CF delivery for patients that fail to trigger equipment
(e.g. mouth breather with closed soft palate.)
If the above remedies do not work, contact DeVilbiss.
Cannot change the flow rate. 1. Flow rate is locked.
2. Ribbon connector cable is disconnected.
3. Unit malfunction.
1. Unlock flow rate.
2. Reconnect cable to circuit board.
2. Contact DeVilbiss.
Oxygen level out of
specification.
Intake silencer/filter is blocked. Replace intake silencer/filter. If the above remedies do not work,
contact DeVilbiss.
Yellow Low Oxygen light
illuminating.
OR
Red Service Required
illuminating and audible alert
sounding. External Power and/
or Battery Power lights
illuminating. Unit is operating.
1. Air filter is blocked.
2. Exhaust is blocked.
3. Intake silencer/filter is blocked.
1. Check the air filter. If the filter is dirty, wash it following the
cleaning instructions.
2. Check the exhaust area; make sure there is nothing
restricting the unit exhaust.
3. Replace intake silencer/filter.
If the above remedies do not work, contact DeVilbiss.
Red Service Required light
illuminating. External Power
and/or Battery Power lights
illuminating. Audible alert
sounding. Unit not operating.
1. Air filter is blocked.
2. Exhaust is blocked.
3. Unit is overheated.
4. Intake silencer/filter is blocked.
1. Check the air filter. If the filter is dirty, wash it following the
cleaning instructions.
2. Check the exhaust area; make sure there is nothing
restricting the unit exhaust.
3a. Allow unit to cool and try again.
3b. Move unit to cooler location.
4. Replace intake silencer/filter.
If the above remedies do not work, contact DeVilbiss.
Power Fail alert activated:
External Power light flashing
and audible alert sounding.
1. Battery completely discharged.
2. Lost external power without battery
installed.
1. Recharge battery.
2. Install battery or plug into external power.
Red Service Required light
flashing. Flow. Indicator
light flashing. Audible alert
sounding. Unit operating.
1. Blocked or defective cannula or oxygen
tubing.
2. Tubing inside unit is kinked.
1a. Detach cannula. If proper flow is restored, clean cannula or
replace if necessary.
1b. Disconnect the oxygen tubing at the oxygen outlet. If
proper flow is restored, check oxygen tubing for
obstructions or kinks. Replace if necessary.
2. Unkink tubing.
Unit not operating when used
with charged battery. Power
Fail alert activated: External
Power light flashing and
audible alert sounding.
Battery is overheating. Allow battery to cool.
tRoubleshootInG
LT-2044 Rev A 306 iGo SM.indd 9 11/6/09 6:23:22 AM